廠牌:TOMY 日本
供應商:智勤企業有限公司(超低溫冷凍櫃/生物安全櫃/離心機/滅菌鍋/CO2,O2培養箱/化學工作台)
聯絡電話:02-23955078 07-3451553
快速詢價 |
LCX-200
LCX-100
Initial values for acceleration and deceleration characteristics are set from 0 to 500rpm, however, they can be varied up to maximum speed in 100rpm increments by function settings.
LCX-200/100, with the ability to accommodate microtubes to 250ml bottles or deep well plates, are ideal for wide range of applications. Using an optional bucket cap kit B433 *, various types of rotors including swinging bucket rotors of biohazard-proof can be mounted.
Convenient flashing function allows operation with preset centrifuging conditions.
*Bucket cap kit B433 is an optional kit.
This highly sealed rotor, which has been tested and certified by the Health Protection Agency (HPA, UK) to be compliant with the International Standard (Annex AA of the IEC 61010-2-020), meets requirements for biohazard safety. [Bucket cap kit B433] is an optional kit.
Easy-to-see LCD monitor display shows various information including remaining run time, error messages, rotor types in use and records of centrifugal operation.
*Indication displays on the control panel in the photo is different from displays of actual operation.
Dual color LED display allows the user to monitor the operation status just by looking the display; red indicates that the rotor is spinning while green indicates the rotor is stopped.
Innovative Flex-Spin system employed to maintain safe and efficient centrifugal operation allows a margin of error and eye-balancing of sample tubes.
*Some rotors are excepted.
• LCX-200/100 are equipped with a function to automatically reduce the rotation speed and shut the system down if it is being run over the imbalance tolerance limit.
The Flex-Spin Drive features a flexible motor shaft that equalizes minor imbalances without damaging the centrifuge or triggering the imbalance protection system. This allows excellent tolerance for minor differences in sample volumes.
With a single knob push, the rotor can be quickly and securely fixed.
*Some rotors are excepted.
The hand guard device is equipped to prevent operator’s hand and fingers from getting caught by the chamber lid.
The operator can open / close the lid and load / unload the sample tubes even though during the main power switch off.
LCX-Series equip common shaft with other TOMY floor type centrifuges. Various compatible rotors can be used as following centrifuges.
CAX-571
AX-521/511/501
NIX-521
※ Some of combinations of centrifuge, rotor, bucket and rack are not compatible.
Model | LCX-200 | LCX-100 |
Maximum Speed | 10,000rpm | |
Maximum RCF | 11,740G | |
Maximum Capacity | 1,000ml (250ml x 4) | |
Speed Control System | Microprocessor feedback control | |
Drive Motor | Induction motor | |
Drive System | Flex-Spin drive system (Direct drive with an automatic alignment function) | |
Data Entry | Jog Dial equipped with an button | |
Speed Setting Range | 100rpm to 10,000 rpm(100rpm increments) | |
RCF Setting Range | 10 G to 11,740 G (Less than 300G : 10G increments, Over 300G : 100G increments) |
|
Time Setting Range | From 10 sec to 50 sec, in 10 sec increments, from 1 min to 99 min, in 1 min increments or for free |
|
Safety Devices | • Imbalance detector • Lid interlock • Lid open/close detector • Abnormal speed detector • Over-current circuit breaker • Rotor identification system • Motor over-current detector | |
Additional Functions | • Three-step acceleration characteristics selection • Three-step deceleration characteristics selection • Three memory function • FLASH (momentary spin) function • BART Code selection • Memory function for the last run parameters | |
Rated Current | 120V : 10A | - |
220/230/240V : 6A | 220/230/240V : 4A | |
Power Requirements | 1-Phase AC120V 60Hz 15A | - |
1-Phase AC220/230/240V 50/60Hz 8A | 1-Phase AC220/230/240V 50/60Hz 8A | |
Power Consumption(Heat output) | 660W (568kcal/h) | 430W (370kcal/h) |
Breaker | Rated breaking current 120V : 15A | - |
Rated breaking current 220/230/240V : 7.5A | Rated breaking current 220/230/240V : 7.5A | |
Protection Against Electric Shock | Class I | |
Dimensions(mm) | 460W x 584D x 355H (excluding hinge) | 430W x 554D x 355H (excluding hinge) |
Net Weight | 41kg | 39kg |
Environmental Requirements | Ambient temperature range:6 to 35℃, Relative humidity:not exceeding 75%RH, Altitude: up to 2000m, Intended exclusively for use indoors, overvoltage category:II, pollution degree:2, inclination:not exceeding 2 degrees |
|
Accessories Included | • Operator's manual x1 • Warranty Card x1 • BART Code table x1 • Clear storage case for operator's manual x1 • Rotor locking knob x1 • Power cord x1 |
* Specifications and design are subject to change for improvement without prior notice. Please ask for the latest specifications prior to your order and procurement.
*The above specifications show the maximum value of speed, RCF(G) and sample capacity. These values may change according to the combination of rotor, bucket, adaptor and tubes.
連結原廠
http://www.digital-biology.co.jp/manufactured/products/lcx/
Angle Rotors for LCX-200/100
http://www.digital-biology.co.jp/manufactured/products/lcx/pdf/201606_LCX_acce_angle.pdf
Swing Out Rotors for LCX-200/100
http://www.digital-biology.co.jp/manufactured/products/lcx/pdf/201606_LCX_acce_swing.pdf