攜帶式、耐用、低維護的M5600高溫高壓耐酸流變儀是Grace的專利產品,也是高溫高壓耐酸流變儀市場上的一款超值產品。便利實用的使用者軟體大幅縮短了測試時間。儀器集合了多項先進的專利技術,能夠顯著減少停機時間和維護成本。
M5600高溫高壓耐酸流變儀帶有強大的使用者導向的計電腦軟體,支援長時間的酸性測試。與同類產品相比,M5600可實現複雜、動態和振蕩的高溫高壓耐酸的試驗,可節省最多50%的試驗時間。堅固的結構設計使M5600適用於更苛刻的測試環境。
M5600新添加的動態振蕩測量功能,用戶可以很容易地將API測量與動態振蕩測量進行關聯,使相關測試一次完成。它開創了一個油田流體、固相控制和油氣增產測試的新領域。非常簡單易學的操作軟體,使其既可用於複雜的研究又可以進行產品的品質控制。
優勢:
特點:
規格:
尺寸: |
8.5" x 12.5" x 25.5" |
重量: |
66 lbs (附油槽) |
關鍵結構: |
哈氏合金 C (潤濕部分) |
樣品體積: |
32 ~ 78 ml |
溫度範圍: |
環境溫度 (20 °F 配冷卻器) ~ 500 °F |
速度範圍: |
0.0001 ~ 1,100 rpm 連續 |
黏度範圍: |
0.5 ~ 5,000,000 Cp |
壓力範圍: |
大氣壓~ 1000 psi |
剪切速率: |
0.00004 ~ 1870 sec-1 |
頻率: |
0.01 ~ 5 Hz (黏彈性選項) |
振幅: |
0.1% ~ 500 % (黏彈性選項) |
精度: |
0.01% 全量程 |
重複性: |
±0.05% 全量程 |
力矩範圍: |
14 μN.m ~ 100 mN.m |
剪切應力: |
1 ~ 15,000 dyne/cm2 |
電源電壓: |
120V 或240V (配置變壓器) |
電源頻率: |
50 Hz或60 Hz |
電腦: |
Windows 9X/2000/XP/Vista |
Description
Rotational viscometer with direct-coupled torque transducer - no bearings
Due to the patented design, The M5600 HPHT Rheometer provides direct reading inside the pressure vessel and employs no bob shaft bearings. That means lower maintenance costs while allowing continuous testing of corrosive samples.
The only true rheology measurement under pressure
The M5600 hardware design incorporates a direct drive between the bob shaft and the torque transducer, which eliminates momentum of inertia errors associated with magnetically coupled torque transducers. This, due to the elimination of bob shaft bearings, allows the torque transducer to respond quickly and consistently to changing bob shaft torque.
Viscoelastic option available for G', G" and phase angle testing
The M5600 HPHT Rheometer does not employ magnetic coupling, instead taking direct measurements inside the pressure vessel, providing true measurement in a pressurized environment. This allows us to offer the M5600 dynamic option, which enables the user to perform true G' and G" viscoelasticity tests under pressure, without having to pre-test using fluids with known G'/G" values.
Cutting-edge database software enables customized search and data comparison
The application software, M5600 PC, includes powerful database tools for searching, categorizing, and comparing test results. The search criteria can include any test result parameter, including additives, descriptions, or other details. M5600 PC also allows you to overlap as many test charts as you like. Each test chart will display with customized graphics to differentiate it from other tests
Patented design & detailed engineering lead to highly consistent data
The unique hardware and software design of the M5600 enable it to produce test results that are extremely repeatable, because process errors and other complications are eliminated by the robust and innovative configuration of hardware and software. By removing contamination and other interference from the test environment, the M5600 test results reflect the properties of the test fluid instead of the properties of the testing hardware.
The Grace Instrument M5600 HPHT Rheometer is a true Couette, coaxial cylinder, rotational, high pressure and temperature rheometer (up to 2,000 psi and 500°F).
The M5600 HPHT Rheometer is also available with an optional viscoelastic module for performing viscoelasticity tests to derive G', G”, G* and phase angle. This vastly increases the researcher's ability to predict the behavior of fluids, such as their capacity for suspending and carrying solids (weight material sag, drill cuttings transport, proppant transfer, etc.)
In the field of rheology-measuring equipment, only the Grace Instrument M5600 HPHT Rheometer offers true visco-elasticity measurement under pressure. Due to the patented design, The M5600 HPHT Rheometer provides direct reading inside the pressure vessel and employs no bob shaft bearings. That means lower maintenance costs while allowing continuous testing of corrosive samples.
Other rheometers often utilize a magnetic coupling across a pressure vessel to rotate a magnet on a bob shaft, using metal or air bearings, in order to measure viscoelasticity. A calibration fluid with known G'/G" values must be tested to provide a baseline for viscoelastic measurement. However, when the tested fluid possesses different G'/G" values compared to the calibration fluid, G'/G" measurement will be invalid.
The M5600 HPHT Rheometer does not employ magnetic coupling, instead taking direct measurements inside the pressure vessel, providing true measurement in a pressurized environment. This allows the user to perform true G' and G" viscoelasticity tests under pressure, without having to pre-test using fluids with known G'/G" values.
The M5600 HPHT Rheometer incorporates a direct drive between the bob shaft and the torque transducer, which eliminates momentum of inertia errors associated with magnetically coupled torque transducers. This, due to the elimination of bob shaft bearings, allows the torque transducer to respond quickly and consistently to changing bob shaft torque.
Features
Specifications
U.S. Patents: 6,938,464 U.S. Patents: 6,951,127 U.S. Patents: 9,157,846 |
Dimensions: 8.5" long x 12.5" wide x 25.5" tall |
Weight: 66 lbs (with bath) or 61 lbs (with carbon block) |
Pressure Range:: Atm to 2,000 psi |
Sample Size: 32 to 78 ml |
Resolution: 0.01% of full scale range or better |
Temperature Range: Ambient (20 °F with chilling device) to 500 °F |
Speed Range: 0.0001 to 1,100 rpm continuous |
Frequency Range: 0.01 to 5 Hz (Optimized at 0.2 to 3 Hz) |
Amplitude Range: 0.1% to 500 % (Optimized at 0.2% to 500%) |
Shear Rate Range: 0.00004 to 1870 sec-1 |
Shear Stress Range: 1 to 15,000 dyn/cm2 |
Viscosity Range: 0.5 to 5,000,000 Centipoise |
Torque: 14 μN.m to 100 mN.m |
Repeatability: ±0.05% of full scale range or better |
Power Voltage: 120V or 240V (with transformer) |
Construction: Hastelloy C wetted material |
Power Frequency: 50 Hz or 60 Hz |
Computer Requirements: PC with MS Windows 9X/2000/XP/Vista/Win7 |