廠牌:日東精工分析科技
供應商:南北貿易有限公司 日東精工分析科技(三菱化學分析科技) 台灣總代理
聯絡電話:04-22994269
快速詢價 |
商品規格 | |
商品名稱: |
卡式微量水份計 CA-310型 (庫倫法) |
特點: |
♦ 擴充功能與多種的搭配應用,最大4頻道同時測量 |
螢幕:
|
8.4英吋彩色液晶觸控式螢幕 |
重量:
|
主機 CA-310MC: 約 2kg 攪拌座 CA-310STR: 約 2.5kg |
尺寸:
|
主機 CA-310MC: 約 245 W× 160 D× 215 Hmm 攪拌座 CA-310STR: 約 120 W× 340 D× 135 Hmm |
請選用三菱水份儀專用之 陽陰極液、滴定液、脫水溶劑及標準品…等試劑
選購品 |
||||||||
|
||||||||
|
|
|||||||
|
||||||||
|
|
|||||||
|
||||||||
|
|
|||||||
|
||||||||
|
|
|||||||
|
||||||||
|
|
|||||||
|
||||||||
|
|
|||||||
|
||||||||
卡氏水份計;卡氏微量水份計;水份計;水份儀;水份測定儀;水分測定儀;水分分析儀;水份分析儀;卡式微量水分計;微量水分儀;微庫倫水分測定器;微庫倫水份測定器;分析儀;三菱化學;水分測定器;三菱化學台灣總代理;水份測定器;CA-310;Karl-Fischer Moisture Meter;Mitsubishi Chemical Analytech Co Ltd;卡氏水分計;卡氏水分儀;卡氏水份測定儀;微量水份分析儀;微量水份計;卡式水份儀;三菱化學分析科技-台灣總代理;水分計;水份計;水分測定器;水份測定器;水分測定儀;水份測定儀;水分分析儀;水份分析儀;卡氏水分計;卡氏水份計;卡式水分計;卡式水份計;卡氏水分儀;卡氏水份儀;卡式水分儀;卡式水份儀;卡氏水分測定儀;卡氏水份測定儀;卡式水分測定儀;卡式水份測定儀;微量水分計;微量水份計;微量水分分析儀;微量水份分析儀;Karl Fischer Moisture Titrator;
什麼是卡爾費歇爾法?
如下面化學反應公式(1)所示,卡爾‧費歇爾法使用卡爾‧費歇爾試劑測定水分含量,該試劑定量地、有選擇性地與水發生反應。卡爾‧費歇爾試劑的組成成分是碘、二氧化硫、鹽基和醇類等溶劑。
I2 + SO2 + 3 Base + ROH + H2O → 2 Base‧HI + Base‧HSO4R…….(1)
如下所述,這種方法可以用於容量及庫倫滴定系統。
圖1:
庫倫測定庫倫滴定
用庫倫滴定法滴定。將試樣加入電解液中,其主要成分是碘化物離子、二氧化硫、鹽基以及醇類等溶劑。如公式(2)所示,通過電解氧化作月產生碘,直接與卡爾‧費歇爾試劑發生反應。
2I--2e→I2…….(2)
按照法拉第定律,碘的產生與電量成正比。這就意味著,含水量可以從電解氧化所需庫倫數直接得出來。
1mg的水=10.71庫倫
如圖1所示,需要兩種類型的庫倫試劑:一種是陽極電解液,置電解槽的陽極室中,另外一種是陰極電解液,置於電解槽的陰極室中。庫倫試劑不需要標定。另一個優點是可以反複使用。
容積滴定容量滴定
在一個燒中放入適合於試樣的脫水溶劑。用滴定劑從該溶劑中轉移所有的水分。然後,加入試樣。用滴定劑進行滴定,滴定量(mgH2O/mL)是事先生確定的。試樣的水分含量由滴定容量(mL)確定的。使用恒定電流極化電壓法檢測終點。圖2顯示了典型的在市場上銷售的自動容積滴定系統的各個組成部分。
AQUAMICRON®是三菱化學株式會社註冊商標。
“日東精工分析科技(前身為三菱化學)集團的子公司API公司”於2002年10月開始成力地開展KF(卡爾‧費歇爾)試劑業務。向客戶提供性能卓越的試劑、服務和技術支援,令客戶滿意。
KF(卡爾‧費歇爾)試劑-AQUAMICRON在三菱化學的鹿島工廠生廠,工廠的質量系統達到ISO 9001的要求,並得到了第三方認證。
CA-310《Coulometric》 | ||
Highly scalable with maximum 4-channel simultaneous measurement controlled by wide color touch screen. |
||
CA-310 Specifications 《Coulometric Moisture Meter》 | |
|
|
Measurement method:
|
Coulometric Karl-Fischer titration, 4-channel simultaneous measurement (option) |
|
|
Titration control: | Constant Current pulse timer control |
|
|
End point detection: | Constant current polarization potential |
Electrolysis current: | 430mA (fast dehydration mode 516mA) |
|
|
Titration speed: | Avg. 2.2mgH2O /min (36μgH2O/sec.) |
|
|
Background: | Automatic correction, constant display, background level at start of measurement is displayed |
|
|
Measuring range: | 5μg - 999.9999mg H2O (5μg measurement: under dry atmosphere) |
|
|
Detection sensitivity: | 0.1μg H2O |
|
|
Accuracy: | <0.3% RSD at 1mg or more H2O |
|
|
Stirring method: | Magnetic stirrer |
|
|
Titration cell: | Closed, Capacity 150mL |
|
|
Display: | 8.4 inch color LCD Touch screen |
|
|
File memory: | Sample parameters: 99 Schedules: 99 Results: 9,999 |
|
|
Calculation functions: | Concentration calculation, statistical calculation, recalculation, reanalysis of titration curve |
|
|
Printer: | Option, thermal or dot impact |
|
|
Vaporizer operating functions: |
VA-300, VA-200, VA-210, VA-230, VA-236S, VA-121, VA-122, VG-200 (stepped temperature control available for each vaporizer) |
|
|
External I/O: | Electronic balance: Automatic input of weight USB: 4 ports LAN: 1 port |
|
|
Additional functions: | Data Integrity (GLP/GMP compliance), detection electrode conditioning |
|
|
Operating conditions: | Temperature: 15-40oC Relative humidity: below 85% (without condensation) |
|
|
Dimensions, Weight: | Main Unit: Approx. 245(W)× 160(D)× 215(H) mm, 2kg Stirrer: Approx. 120(W)× 340(D)× 135(H) mm, 2.5kg |
|
|
Options |
|
Sample Boat Type model VA-300 Various types of plastics, rubber, foods, pharmaceuticals, inorganic salts and other powders and solid samples Heating temperature: 50~300oC |
|
Disposable Vial Type Heating temperature: 70~300oC |
|
Autosamplers and Vaporizers Heating temperature: 70~300oC |
|
Autosamplers and Vaporizers Applicable gases: Propane. Butane,other liquefide gases(Up to 5 different gas factors can be set) |
|
High Temperatures Iron ore, clay, metal oxides, inorganic salts Heating temperature: 250~1000oC |
|
High/Low Temperatures Enables measurement of ancillary moisture. Iron ore, clay, welding rods, metal oxides, inorganic salts Heating temperature: 70~300oC, 250~1000oC |
|